SEC-elite


The SEC-elite is a programmable multi-line gas proving and detection system capable of pressure proving 4 gas lines and monitoring up to 8 detectors from a single panel.

With each gas line controlled individually, laboratories with multiple gas lines no longer need several panels installed. This reduces installation and purchase costs and creates a simple interaction for the user, which is aided by hidden until lit switches.

   

Description

Detection of exotic gases can also be monitored.

All features of the system can be enabled during the initial installation, or they can be introduced at a later date, thereby providing a unique level of future-proofing.

Service Isolation

Isolation control of electrics, water valves and non-proving gas valves can also be achieved.

Technical Details

Product Dimensions 212mm x 183mm x 97mm
Gas Detection The SEC has been designed to accept multiple gas detectors, internal programmable relay options can provide demand-controlled ventilation, BMS signals and isolation of gas valves upon detection of the target gases at prescribed levels. The current range is:
Methane/Natural Gas (CH4), Propane (C3H8), Carbon Monoxide (CO), Carbon Dioxide (CO2), Oxygen – enrichment and depletion. (O2), Hydrogen (H)Coming soon will be Ammonia (NH3), Acetylene (C2H2), Nitrogen Dioxide (NO2), Chlorine (CL) detection.
Alarm Types Each detector can be set to either a peak alarm or time-weighted average (TWA) and be set to either non-latching or latching with a programmable time to latch of between 1 & 59 seconds.
Peak Alarms Can be set to any value with the particular gas detectors range (during the setup and commissioning by Medem engineers).
• Time Weighted Average (TWA) Alarms
• Three different TWA alarms can be set
(e.g. 5000ppm 10 secs, 3000ppm 2mins, 500ppm @10mins.)
Logging There is a downloadable detector log of minute average data for each detector with up to 90 days history.
Warranty Medem’s industry-leading warranties are a demonstration of the company’s confidence in the quality and design of our products.
The system will be pre-configured before supply but will require full commissioning by a Medem engineer, following the commissioning the main panel will be issued with a 10-year warranty.At the time of commissioning, photos and gas level readings are recorded along with other relevant details of the installation. A 24-hour helpline number and the serial number is attached with the panel and all details are then entered on to the Medem database.
Gas detectors carry a two-year warranty.
Service The panel and sensors must be re-commissioned and tested every 12 months as a minimum but with consideration of every 6 months dependant on gas type and environment. This isn’t just a bump test but the application of calibrated gas to the sensors and using a Medem ‘calim board’ to either calibrate or confirm sensor accuracy.

Downloads


File File size
pdf SEC-elite v1.1 Instructions mp 1 MB